Serial Optical Coherence Tomography at Baseline, 7 Days, and 1, 3, 6 and 12 Months After Bioresorbable Scaffold Implantation in a Growing Porcine Model

Circ J. 2019 Feb 25;83(3):556-566. doi: 10.1253/circj.CJ-18-0855. Epub 2019 Jan 31.

Abstract

Background: Little is known about serial changes in lumen and device dimensions after bioresorbable scaffold implantation in a growing animal model. Methods and Results: ABSORB (n=14) or bare metal stents (ICROS amg [Abbott Vascular, Santa Clara, CA, USA], Winsen-Luhe, Germany; n=15) were implanted in the coronary arteries of domestic swine (a hybrid of Finnish-Norwegian Landrace swine) weighing 30-35 kg. Angiography and optical coherence tomography (OCT) were performed immediately after implantation and repeated at 7 days, 1, 3, 6 and 12 months after the index procedure. One month after implantation, mean lumen area decreased relative to baseline in both groups (relative area change from baseline, -41.4±15.6% for ABSORB vs. -20.9±18.6% for ICROS) while mean device area decreased only in the ABSORB group (relative area change: -11.1±9.4% vs. +0.14±7.95%, respectively). At 12 months, mean lumen area increased relative to baseline in both groups (relative area change from baseline, +55.6±22.4% vs. +32.3±83.6%, respectively) in accordance with the swine growth weighing up to 260-300 kg. Mean device area in the ICROS group remained stable whereas that in the ABSORB group began to increase between 3 and 6 months along with the vessel growth (relative area change: +107.8±25.7% vs. +0.14±7.95%).

Conclusions: In the growing porcine model, ABSORB was associated with greater extent of recoil 1 month after implantation compared with ICROS but demonstrated substantial adaptability to vessel growth in late phase.

Keywords: Coronary artery disease; Pediatrics; Percutaneous coronary intervention; Stent.

MeSH terms

  • Absorbable Implants / standards*
  • Animals
  • Coronary Angiography / methods
  • Coronary Vessels / diagnostic imaging*
  • Coronary Vessels / growth & development
  • Coronary Vessels / surgery
  • Models, Animal
  • Prosthesis Design / standards
  • Stents / standards*
  • Swine
  • Time Factors
  • Tomography, Optical Coherence / methods*