[Rapid on-site screening of five prohibited ingredients in cosmetics using thermal desorption-corona discharge ionization coupled with ion mobility spectrometry]

Se Pu. 2019 Feb 8;37(2):233-238. doi: 10.3724/SP.J.1123.2018.09005.
[Article in Chinese]

Abstract

A rapid on-site analytical method is developed for screening lidocaine and four other types of prohibited ingredients in cosmetics using thermal desorption-corona discharge ionization together with ion mobility spectrometry. Performing no pretreatment, we dropped or sprayed cosmetic samples onto a Nomex sampling swab. The sampler was placed into a compartment for thermal desorption and corona discharge ionization; then, ion mobility spectrometry analysis was performed. The limits of detection (LODs) for the five analytes ranged from 10 to 50 ng. The instrumental analysis time for a single run was less than 20 ms, and the total sample analysis period was within 1 min. The proposed method is simple, fast, and has low cost, and could be used as an analytical tool for rapid on-site screening of prohibited ingredients in cosmetics.

Keywords: cosmetics; ion mobility spectrometry; rapid on-site screening; thermal desorption.

MeSH terms

  • Cosmetics / chemistry*
  • Ion Mobility Spectrometry*
  • Lidocaine / analysis*
  • Limit of Detection