A Model to Study Thermal Energy Delivery to the Choroid: A Comparison of Surgical Devices

Transl Vis Sci Technol. 2018 Dec 28;7(6):39. doi: 10.1167/tvst.7.6.39. eCollection 2018 Nov.

Abstract

Purpose: We measure and compare surgical devices using an ex vivo, temperature-controlled, choroidal incision model during thermal energy transfer with a high-resolution infrared camera.

Methods: Ex vivo porcine choroidal tissue specimens (n = 516) were isolated and placed on a temperature-regulated (37°C) perfusion platform. We tested the pulsed electron avalanche knife (PEAK), micropulse laser (MpL), continuous laser (CL), and bipolar cautery (BpC) at three energy settings (11 [low], 45 [medium], and 134 [high] mJ/mm). Each device was clamped to a stationary mechanical arm. Movement of tissue specimens beneath the surgical device was achieved using a stepping motor-driven x-y table. An infrared video camera measured orthogonal temperature variation in the surrounding tissue.

Results: Increased power resulted in greater lateral thermal spread using all modalities (P < 0.001). Mean (standard deviation) lateral thermal spread at low energy was smallest for the MpL at 0.0 (0.01) mm (P < 0.001), whereas BpC had the least collateral tissue damage at medium and high energies (0.02 [0.08] and 0.34 [0.22] mm, respectively; P < 0.001). Fluidics of the ex vivo system may limit thermal spread. The PEAK had the greatest thermal spread across all energy groups (P < 0.001), with clinically relevant variation between disposable blades.

Conclusions: Our ex vivo model enabled direct comparison of threshold thermal tissue injury across four devices. MpL and BpC showed the least thermal damage. PEAK had a higher variation in energy delivery, but also has the advantage of more effective tissue cutting.

Translational relevance: Our ex vivo surgical device analysis provides thermal tissue injury predictions for choroidal surgery.

Keywords: PEAK; choroid; lateral thermal spread; micropulse laser; retinal pigment epithelium.