Assessment of diabetic retinopathy using two ultra-wide-field fundus imaging systems, the Clarus® and Optos™ systems

BMC Ophthalmol. 2018 Dec 20;18(1):332. doi: 10.1186/s12886-018-1011-z.

Abstract

Background: The ability to image wide fundus fields and to conduct swift, non-invasive examinations is increasingly important with the escalation in patients with diabetic retinopathy (DR).

Methods: Fifty eyes of 28 consecutive patients with DR were examined in this prospective observational study. A total of 46 eyes, 25 right and 21 left eyes, of 27 patients (male, 19; female, 8) were ultimately included in the analysis. All patients underwent comprehensive ophthalmological examination. A single image each was obtained using two ultra-wide-field (UWF) imaging systems: Optos® (Optos Carfornia®, Optos PLC, Dunfermline, United Kingdom) and Clarus™ (CLARUS 500™, Carl Zeiss Meditec Inc., Californea, USA), without mydriasis. The total retinal area captured and the obscured retinal area were compared between the two systems using nonparametric Wilcoxon matched-pairs signed-rank analysis. Early Treatment of Diabetic Retinopathy Study (ETDRS) and International Clinical DR severity were analyzed by κ statistics.

Results: The Optos® allowed capture of larger areas of the fundus than the Clarus™ (465 ± 117 vs. 243 ± 39 disc areas, P < 0.0001). In 85% (39/46) of Optos® images and 7% (3/46) of Clarus™ images, a slightly obscured area was observed within the ETDRS-7 field area. κ values for ETDRS DR severity and International Clinical DR severity between the Optos® and Clarus™ images were 0.88 and 0.79, respectively. Severity was higher according to Clarus™ images in two eyes in which the ETDRS DR severity grading differed between the systems. Severity was higher in four Clarus™ images and in a single Optos® image in five eyes in which the International Clinical DR severity grading differed between the systems.

Conclusion: The Optos® and Clarus™ UWF retinal imaging systems were useful for examining eyes with DR, using single images obtained without mydriasis. The systems were both generally consistent in assessing DR severity, with some partial discrepancies. It is important to understand the characteristics of each respective UWF retinal imaging system when using them to assess DR.

Keywords: Diabetic mellitus; Diabetic retinopathy; Early treatment of diabetic retinopathy study diabetic retinopathy severity; International clinical diabetic retinopathy severity; Ultra-wide-field retinal imaging systems.

Publication types

  • Observational Study

MeSH terms

  • Adult
  • Aged
  • Diabetic Retinopathy / diagnostic imaging*
  • Diagnostic Techniques, Ophthalmological*
  • Female
  • Fundus Oculi*
  • Humans
  • Male
  • Middle Aged
  • Photography / methods
  • Prospective Studies
  • Retina / diagnostic imaging*
  • Retina / pathology
  • Sensitivity and Specificity