Multi-institutional comparison of secondary check of treatment planning using computer-based independent dose calculation for non-C-arm linear accelerators

Phys Med. 2018 Dec:56:58-65. doi: 10.1016/j.ejmp.2018.11.011. Epub 2018 Nov 23.

Abstract

Purpose: This report covers the first multi-institutional study of independent monitor unit (MU)/dose calculation verification for the CyberKnife, Vero4DRT, and TomoTherapy radiotherapy delivery systems.

Methods: A total of 973 clinical treatment plans were collected from 12 institutions. Commercial software employing the Clarkson algorithm was used for verification after a measurement validation study, and the doses from the treatment planning systems (TPSs) and verification programs were compared on the basis of the mean value ± two standard deviations. The impact of heterogeneous conditions was assessed in two types of sites: non-lung and lung.

Results: The dose difference for all locations was 0.5 ± 7.2%. There was a statistically significant difference (P < 0.01) in dose difference between non-lung (-0.3 ± 4.4%) and lung sites (3.5 ± 6.7%). Inter-institutional comparisons showed that various systematic differences were associated with the proportion of different treatment sites and heterogeneity correction.

Conclusions: This multi-institutional comparison should help to determine the departmental action levels for CyberKnife, Vero4DRT, and TomoTherapy, as patient populations and treatment sites may vary between the modalities. An action level of ±5% could be considered for intensity-modulated radiation therapy (IMRT), non-IMRT, and volumetric modulated arc radiotherapy using these modalities in homogenous and heterogeneous conditions with a large treatment field applied to a large region of homogeneous media. There were larger systematic differences in heterogeneous conditions with a small treatment field because of differences in heterogeneity correction with the different dose calculation algorithms of the primary TPS and verification program.

Keywords: Independent calculation verification; Institution-specific tolerance; Multi-institutional comparison; Quality assurance; non-C-arm linear accelerator.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Algorithms
  • Humans
  • Lung
  • Particle Accelerators*
  • Quality Assurance, Health Care* / methods
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted* / methods
  • Radiotherapy, Intensity-Modulated / instrumentation
  • Retrospective Studies