Field suitability and diagnostic accuracy of the Biocentric® open real-time PCR platform for plasma-based HIV viral load quantification in Swaziland

BMC Infect Dis. 2018 Nov 14;18(1):570. doi: 10.1186/s12879-018-3474-1.

Abstract

Background: Viral load (VL) testing is being scaled up in resource-limited settings. However, not all commercially available VL testing methods have been evaluated under field conditions. This study is one of a few to evaluate the Biocentric platform for VL quantification in routine practice in Sub-Saharan Africa.

Methods: Venous blood specimens were obtained from patients eligible for VL testing at two health facilities in Swaziland from October 2016 to March 2017. Samples were centrifuged at two laboratories (LAB-1, LAB-2) to obtain paired plasma specimens for VL quantification with the national reference method and on the Biocentric platform. Agreement (correlation, Bland-Altman) and accuracy (sensitivity, specificity) indicators were calculated at the VL thresholds of 416 (2.62 log10) and 1000 (3.0 log10) copies/mL. Leftover samples from patients with discordant VL results were re-quantified and accuracy indicators recalculated. Logistic regression was used to compare laboratory performance.

Results: A total of 364 paired plasma samples (LAB-1: n = 198; LAB-2: n = 166) were successfully tested using both methods. The correlation was high (R = 0.82, p < 0.01), and the Bland-Altman analysis showed a minimal mean difference (- 0.03 log10 copies/mL; 95% CI: -1.15 to 1.08). At the clinical threshold level of 3.0 log10 copies/mL, the sensitivity was 88.6% (95% CI: 78.7 to 94.9) and the specificity was 98.3% (95% CI: 96.1 to 99.4). Sensitivity was higher in LAB-1 (100%; 95% CI: 71.5 to 100) than in LAB-2 (86.4%; 95% CI: 75.0 to 94.0). Most upward (n = 8, 2.2%) and downward (n = 11, 3.0%) misclassifications occurred at the 2.62 log threshold, with LAB-2 having a 16 (95% CI: 2.26 to 113.27; p = 0.006) times higher odds of downward misclassification. After retesting of discordant leftover samples (n = 17), overall sensitivity increased to 93.5% (95% CI: 85.5 to 97.9) and 97.1% (95% CI: 90.1 to 99.7) at the 2.62 and 3.0 thresholds, and specificity increased to 98.6% (95% CI: 96.5 to 99.6) and 99.0% (95% CI: 97.0 to 99.8) respectively.

Conclusions: The test characteristics of the Biocentric platform were overall comparable to the national reference method for VL quantification. One laboratory tended to misclassify VL results downwards, likely owing to unmet training needs and lack of previous hands-on practice.

Keywords: Accuracy; Biocentric; HIV; Open platform; Swaziland; Viral load.

Publication types

  • Evaluation Study

MeSH terms

  • Adolescent
  • Adult
  • Diagnostic Tests, Routine / economics
  • Diagnostic Tests, Routine / methods
  • Diagnostic Tests, Routine / standards
  • Eswatini
  • Female
  • HIV Infections / blood
  • HIV Infections / diagnosis*
  • HIV Infections / virology
  • HIV-1 / genetics*
  • HIV-1 / isolation & purification
  • Humans
  • Male
  • Middle Aged
  • Point-of-Care Systems / economics
  • Point-of-Care Systems / standards
  • Predictive Value of Tests
  • RNA, Viral / analysis
  • RNA, Viral / blood*
  • Real-Time Polymerase Chain Reaction / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Serologic Tests
  • Specimen Handling / methods
  • Viral Load / genetics
  • Viral Load / methods*
  • Young Adult

Substances

  • RNA, Viral