Biomechanical Considerations in Foot and Ankle Circular External Fixation: Maintenance of Wire Tension

Clin Podiatr Med Surg. 2018 Oct;35(4):443-455. doi: 10.1016/j.cpm.2018.05.004. Epub 2018 Aug 11.

Abstract

Initial tensioning of the forefoot wires to 130 kg followed by simultaneous tensioning of the calcaneal wires to 90 kg and using the rigid double-row foot plate closed anteriorly via threaded rods produce maximum preservation of the initial wire tension during foot circular external fixation.

Keywords: Calcaneal wire pretension; Circular external fixation; Foot and ankle; Foot frame biomechanics; Metatarsal wire pretension; Wire tension; Wire tension loss.

Publication types

  • Review

MeSH terms

  • Bone Plates*
  • Bone Wires*
  • Calcaneus / surgery
  • External Fixators*
  • Foot Joints / physiology*
  • Fracture Fixation / instrumentation*
  • Fracture Fixation / methods
  • Humans
  • Metatarsal Bones / surgery