Fabrication of a Urea Biosensor for Real-Time Dynamic Fluid Measurement

Sensors (Basel). 2018 Aug 9;18(8):2607. doi: 10.3390/s18082607.

Abstract

In this study, a portable urea sensor that monitors the urea concentration in flow conditions was fabricated. We propose an electrochemical sensor that continually measures the urea concentration of samples flowing through it at a constant flow rate in real time. For the electrochemical sensing, a porous silk fibroin membrane with immobilized urease was mounted in a polydimethylsiloxane (PDMS) sensor housing. The fabricated urea sensor elicited linear current⁻concentration characteristics in the clinically significant concentration range (0.1⁻20 mM) based on peritoneal dialysis. The sensor maintained the linear current⁻concentration characteristics during operation in flow conditions.

Keywords: amination; flow conditions; real-time monitoring; silk fibroin; urea biosensor.

MeSH terms

  • Biosensing Techniques*
  • Electrochemical Techniques
  • Enzymes, Immobilized / metabolism
  • Fibroins
  • Porosity
  • Rheology*
  • Silk
  • Time Factors
  • Urea / analysis*
  • Urease / metabolism

Substances

  • Enzymes, Immobilized
  • Silk
  • Urea
  • Fibroins
  • Urease