Reduction of the Measurement Time by the Prediction of the Steady-State Response for Quartz Crystal Microbalance Gas Sensors

Sensors (Basel). 2018 Jul 31;18(8):2475. doi: 10.3390/s18082475.

Abstract

This paper presents a new approach to reduce the measurement time by the prediction of the steady-state using the transient response to ethanol for quartz crystal microbalance gas sensors coated with ethyl cellulose. The experimentally measured response curves were successively fitted using a mathematical model based on the sum of two exponentials with different time constants. The parameters of the model were determined, and the time constants and the magnitude of the steady-state response were analyzed. Even though the time constants did not stabilize well, the parameter corresponding to the magnitude of the steady-state response quickly converged and stabilized after 37 s. Moreover, this calculated parameter was highly correlated with the measured values of the steady-state response, which was measured at five times the longest time constant (83 s) of the model. Therefore, the steady-state response could be predicted with a 55% reduction in the measurement (detection) time.

Keywords: QCM gas sensor; measurement time reduction; steady-state response prediction; transient response.