Audible Feedback Improves Internal Model Strength and Performance of Myoelectric Prosthesis Control

Sci Rep. 2018 Jun 4;8(1):8541. doi: 10.1038/s41598-018-26810-w.

Abstract

Myoelectric prosthetic devices are commonly used to help upper limb amputees perform activities of daily living, however amputees still lack the sensory feedback required to facilitate reliable and precise control. Augmented feedback may play an important role in affecting both short-term performance, through real-time regulation, and long-term performance, through the development of stronger internal models. In this work, we investigate the potential tradeoff between controllers that enable better short-term performance and those that provide sufficient feedback to develop a strong internal model. We hypothesize that augmented feedback may be used to mitigate this tradeoff, ultimately improving both short and long-term control. We used psychometric measures to assess the internal model developed while using a filtered myoelectric controller with augmented audio feedback, imitating classification-based control but with augmented regression-based feedback. In addition, we evaluated the short-term performance using a multi degree-of-freedom constrained-time target acquisition task. Results obtained from 24 able-bodied subjects show that an augmented feedback control strategy using audio cues enables the development of a stronger internal model than the filtered control with filtered feedback, and significantly better path efficiency than both raw and filtered control strategies. These results suggest that the use of augmented feedback control strategies may improve both short-term and long-term performance.

MeSH terms

  • Amputees
  • Artificial Limbs*
  • Electromyography
  • Feedback, Sensory*
  • Female
  • Humans
  • Male
  • Models, Biological*
  • Prosthesis Design*
  • Psychometrics