Fully electronic urine dipstick probe for combinatorial detection of inflammatory biomarkers

Future Sci OA. 2018 Mar 27;4(5):FSO301. doi: 10.4155/fsoa-2017-0142. eCollection 2018 Jun.

Abstract

Aim: An electrochemical urine dipstick probe biosensor has been demonstrated using molybdenum electrodes on nanoporous polyamide substrate for the quantitative detection of two inflammatory protein biomarkers, CRP and IL-6.

Materials & methods: The electrode interface was characterized using ζ-potential and Fourier transform infrared spectroscopy. Detection of biomarkers was demonstrated by measuring impedance changes associated with the dose concentrations of the two biomarkers. A proof of feasibility of point-of-care implementation of the biosensor was demonstrated using a portable electronics platform.

Results & conclusion: Limit of detection of 1 pg/ml was achieved for CRP and IL-6 in human urine and synthetic urine buffers. The developed portable hardware demonstrated close correlation with benchtop equipment results.