Adaptive Color Gain for Vena Contracta Quantification in Valvular Regurgitation

Ultrasound Med Biol. 2018 Aug;44(8):1770-1777. doi: 10.1016/j.ultrasmedbio.2018.04.002. Epub 2018 May 18.

Abstract

Severe valvular regurgitation can lead to pulmonary hypertension, atrial fibrillation and heart failure. Vena contracta width is used to estimate the severity of the regurgitation. Parameters affecting visualization of color Doppler have a significant impact on the measurement. We propose a data-driven method for automated adjustment of color gain based on the peak power of the color Doppler signal in the vicinity of the vena contracta. A linear regression model trained on the peak power was used to predict the orifice diameter. According to our study, the color gain should be set to about 6 dB above where color Doppler data completely disappears from the image. Based on our method, orifices with reference diameters of 4, 6.5 and 8.5 mm were estimated with relative diameter errors within 18%, 12% and 14%, respectively.

Keywords: Color Doppler echocardiography; Digital signal processing; Echocardiography; Heart valve diseases.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Echocardiography, Doppler, Color / methods*
  • Heart Ventricles / diagnostic imaging*
  • Heart Ventricles / physiopathology*
  • Mitral Valve Insufficiency / diagnostic imaging*
  • Mitral Valve Insufficiency / physiopathology*
  • Phantoms, Imaging
  • Reproducibility of Results
  • Signal Processing, Computer-Assisted*