X-ray beam-position feedback system with easy-to-use beam-position monitor

J Synchrotron Radiat. 2018 May 1;25(Pt 3):869-873. doi: 10.1107/S1600577518002692. Epub 2018 Mar 14.

Abstract

X-ray beam-position stability is indispensable in cutting-edge experiments using synchrotron radiation. Here, for the first time, a beam-position feedback system is presented that utilizes an easy-to-use X-ray beam-position monitor incorporating a diamond-fluorescence screen. The acceptable range of the monitor is above 500 µm and the feedback system maintains the beam position within 3 µm. In addition to being inexpensive, the system has two key advantages: it works without a scale factor for position calibration, and it has no dependence on X-ray energy, X-ray intensity, beam size or beam shape.

Keywords: X-ray beam-position monitor; diamond-fluorescence screen; feedback system.