Purpose: A minimally invasive, partial-assist, intra-atrial blood pump has been proposed, which would unload the left ventricle with a flow path from the left atrium to the arterial system. Flow modulation is a common strategy for ensuring washout in the pump, but it can increase power consumption because it is typically achieved through motor-speed variation. However, if a pump's performance curve had the proper gradient, flow modulation could be realized passively. To achieve this goal, we propose a pump performance operating curve as an alternative to the more standard operating point.
Methods and results: Mean-line theory was employed to generate an initial set of geometries that were then tested on a hydraulic test rig at ~20,000 r/min. Experimental results show that the intra-atrial blood pump performed below the operating region; however, it was determined that smaller hub diameter and longer chord length bring the performance of the intra-atrial blood pump device closer to the operating curve.
Conclusion: We found that it is possible to shape the pump performance curve for specifically targeted gradients over the operating region through geometric variations inside the pump.
Keywords: Cardiac device; flow modulation; hydraulic design; left ventricular assist device; partial support; rotary blood pump.