Early Experience with a Short, Tapered Titanium Porous Plasma Sprayed Stem with Updated Design

Surg Technol Int. 2018 Jun 1:32:233-238.

Abstract

Introduction: Short stem femoral components in primary total hip arthroplasty (THA) have increased in popularity since the advent of minimally invasive surgical techniques. The concept of a short stem is particularly compatible with tapered designs where the goal is to offload forces proximally in the femur. The purpose of this retrospective review was to review our early experience with a short, tapered titanium femoral component with updated design features.

Materials and methods: Beginning in November 2011 through February 2012, 92 consented patients (93 hips), at a single center, were treated with primary cementless THA using a short stem, tapered femoral component (Taperloc® Complete Microplasty; Zimmer Biomet, Warsaw, Indiana) and were available for review with a minimum two-year follow-up. Mean patient age at surgery was 63.2 years and body mass index (BMI) was 30.8 kg/m2. Mean stem length used was 110.3mm (range, 95-125).

Results: Mean follow-up was 4.5 years (2-6). Harris hip scores improved from 52.5 preoperatively to 84.7 at most recent. One stem was revised the same day for periprosthetic fracture. One patient with early infection was treated with single-stage exchange followed by recurrence that was treated successfully with two-stage exchange. A non-healing wound in one patient was treated with incision and debridement. Radiographic assessment demonstrated no evidence of loosening, osteolysis, distal hypertrophy, or pedestal formation in any hip, and all components appeared well fixed and in appropriate alignment.

Conclusion: In this series of patients treated with primary THA using a short, tapered titanium porous plasma-sprayed femoral component with updated design features, good results were achieved with a low incidence of complications and revision. No aseptic loosening or osteolysis has occurred. Radiographic assessment was excellent for all patients.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Arthroplasty, Replacement, Hip / instrumentation*
  • Female
  • Femur / surgery
  • Hip Joint / surgery
  • Hip Prosthesis*
  • Humans
  • Male
  • Middle Aged
  • Porosity
  • Postoperative Complications
  • Prosthesis Design
  • Retrospective Studies
  • Titanium* / chemistry
  • Titanium* / therapeutic use
  • Treatment Outcome