Factors affecting the separation performance of proteins in capillary electrophoresis

J Chromatogr B Analyt Technol Biomed Life Sci. 2018 Apr 15:1083:63-67. doi: 10.1016/j.jchromb.2018.03.005. Epub 2018 Mar 3.

Abstract

Capillary electrophoresis (CE) is an effective tool for protein separation and analysis. Compared with capillary gel electrophoresis (CGE), non-gel sieving capillary electrophoresis (NGSCE) processes the superiority on operation, repeatability and automaticity. Herein, we investigated the effect of polymer molecular weight and concentration, electric field strength, and the effective length of the capillary on the separation performance of proteins, and find that (1) polymer with high molecular weight and concentration favors the separation of proteins, although concentrated polymer hinders its injection into the channel of the capillary due to its high viscosity. (2) The resolution between the adjacent proteins decreases with the increase of electric field strength. (3) When the effective length of the capillary is long, the separation performance improves at the cost of separation time. (4) 1.4% (w/v) hydroxyethyl cellulose (HEC), 100 V/cm voltage and 12 cm effective length offers the best separation for the proteins with molecular weight from 14,400 Da to 97,400 Da. Finally, we employed the optimal electrophoretic conditions to resolve Lysozyme, Ovalbumin, BSA and their mixtures, and found that they were baseline resolved within 15 min.

Keywords: Capillary electrophoresis; Fluorescence; Protein.

MeSH terms

  • Electrophoresis, Capillary / methods*
  • Proteins / analysis
  • Proteins / chemistry
  • Proteins / isolation & purification*
  • Spectrometry, Fluorescence