Background: In neonatal resuscitation, a ventilation device providing positive end-expiratory pressure (PEEP) is recommended. There is limited information about PEEP delivery in vivo, using different models of self-inflating bag (SIB) at different inflation rates and PEEP settings.
Methods: We compared PEEP delivery to intubated preterm lambs using four commonly available models of paired SIBs and PEEP valves, with a T-piece, with gas flow of 8 L/min. Peak inspiratory pressure inflations of 30 cmH2O, combined with set PEEP of 5, 7 and 10 cmH2O, were delivered at rates of 20, 40 and 60/min. These combinations were repeated without gas flow. We measured mean PEEP, maximum and minimum PEEP, and its difference (PEEP reduction).
Results: A total of 3288 inflations were analysed. The mean PEEP delivered by all SIBs was lower than set PEEP (P<0.001), although some differences were <0.5 cmH2O. In 55% of combinations, the presence of gas flow resulted in increased PEEP delivery (range difference 0.3-2 cmH2O). The mean PEEP was closer to set PEEP with faster inflation rates and higher set PEEPs. The mean (SD) PEEP reduction was 3.9 (1.6), 8.2 (1.8), 2 (0.6) and 1.1 (0.6) cmH2O with the four SIBs, whereas it was 0.5 (0.2) cmH2O with the T-piece.
Conclusions: PEEP delivery with SIBs depends on the set PEEP, inflation rate, device model and gas flow. At recommended inflation rates of 60/min, some devices can deliver PEEP close to the set level, although the reduction in PEEP makes some SIBs potentially less effective for lung recruitment than a T-piece.
Keywords: PEEP delivery; T-piece; low resource setting; neonatal resuscitation; self-inflating bag.
© Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2019. All rights reserved. No commercial use is permitted unless otherwise expressly granted.