Variability induced by the MR imager in dynamic contrast-enhanced imaging of the prostate

Diagn Interv Imaging. 2018 Apr;99(4):255-264. doi: 10.1016/j.diii.2017.12.003. Epub 2018 Feb 7.

Abstract

Purpose: To evaluate the variability induced by the imager in discriminating high-grade (Gleason≥7) prostate cancers (HGC) using dynamic contrast-enhanced MRI.

Material and methods: We retrospectively selected 3T MRIs with temporal resolution<10 seconds and comprising T1 mapping from a prospective radiologic-pathologic database of patients treated by prostatectomy. Ktrans, Kep, Ve and Vp were calculated for each lesion seen on MRI using the Weinmann arterial input function (AIF) and three patient-specific AIFs measured in the right and left iliac arteries in pixels in the center of the lumen (psAIF-ST) or manually selected by two independent readers (psAIF-R1 and psAIF-R2).

Results: A total of 43 patients (mean age, 63.6±4.9 [SD]; range: 48-72 years) with 100 lesions on MRI (55 HGC) were selected. MRIs were performed on imager A (22 patients, 49 lesions) or B (21 patients, 51 lesions) from two different manufacturers. Using the Weinmann AIF, Kep (P=0.005), Ve (P=0.04) and Vp (P=0.01) significantly discriminated HCG. After adjusting on tissue classes, the imager significantly influenced the values of Kep (P=0.049) and Ve (P=0.007). Using patient-specific AIFs, Vp with psAIF-ST (P=0.008) and psAIF-R2 (P=0.04), and Kep with psAIF-R1 (P=0.03) significantly discriminated HGC. After adjusting on tissue classes, types of patient-specific AIF and side of measurement, the imager significantly influenced the values of Ktrans (P=0.0002), Ve (P=0.0072) and Vp (P=0.0003). For all AIFs, the diagnostic value of pharmacokinetic parameters remained unchanged after adjustment on the imager, with stable odds ratios.

Conclusion: The imager induced variability in the absolute values of pharmacokinetic parameters but did not change their diagnostic performance.

Keywords: Dynamic contrast-enhanced imaging; Multiparametric MRI; Perfusion imaging; Prostate cancer.

MeSH terms

  • Aged
  • Contrast Media*
  • Humans
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Prostate / diagnostic imaging
  • Prostatic Neoplasms / diagnostic imaging*
  • Retrospective Studies

Substances

  • Contrast Media