Chemiluminescence determination of melamine with Luminol-K3Fe(CN)6 system

J Pharm Anal. 2011 May;1(2):104-107. doi: 10.1016/S2095-1779(11)70018-1. Epub 2012 Jan 30.

Abstract

A sensitive chemiluminescence (CL) method was developed for determining melamine in urine and plasma samples based on the fact that melamine can remarkably enhance the chemiluminescence of Luminol-K3 Fe(CN)6 system in alkaline medium. The determination conditions were optimized. Under optimum conditions, the chemiluminescence intensity had a good linear relationship with melamine in the range of 9.0 × 10-9 - 7.0 × 10-6 g/mL with a correlation coefficient of 0.9992. The detection limits (3σ) were 3.54 ng/mL for urine sample and 6.58 ng/mL for plasma sample. The average recoveries of melamine were 102.6% for urine sample and 95.1% for plasma sample. Melamine in samples was extracted with liquid-liquid extraction procedures and the assay results coincided very well with that determined with flow injection chemiluminescence method. The method provides a reproducible and stable approach for sensitive detection and quantification of melamine in urine and plasma samples.

Keywords: chemiluminescence; flow injection; melamine; plasma; urine.