A systematic review of adenosine triphosphate as a surrogate for bacterial contamination of duodenoscopes used for endoscopic retrograde cholangiopancreatography

Am J Infect Control. 2018 Jun;46(6):697-705. doi: 10.1016/j.ajic.2017.12.007. Epub 2018 Feb 1.

Abstract

Background: Bacterial culture is the accepted standard to measure the adequacy of high-level disinfection (HLD) of duodenoscopes. Adenosine triphosphate (ATP) bioluminescence assays have been suggested as an alternative method of evaluating the quality of reprocessing. We systematically reviewed published research describing the correlation between ATP and bacterial cultures.

Methods: The primary outcome was the correlation or concordance between concomitantly sampled ATP and bacterial contamination obtained from the instrument channel and/or elevator mechanism of the duodenoscope. A secondary outcome included the reduction in ATP measurements between paired samples before and after stages of duodenoscope reprocessing.

Results: Ten studies were included in the analysis. Four studies reported the relationship between concomitantly sampled ATP and cultures. Three studies reported receiver operating characteristic curves (1 study additionally reported a Wilcoxon rank sum test), and 1 study reported Spearman correlation coefficients and paired dichotomous measurements of ATP and bacterial contamination. All analyses suggested a poor relationship between the 2 measures. Studies measuring ATP before and after manual cleaning and before and after HLD reported a reduction in ATP after the reprocessing stage.

Conclusion: Current research does not support the direct substitution of ATP for bacterial culture surveillance of duodenoscopes. Serial ATP measurement may be a useful tool to evaluate the adequacy of manual cleaning and for training of endoscopic reprocessing staff.

Keywords: Disinfection; ERCP; Reprocessing; Surveillance.

Publication types

  • Comparative Study
  • Evaluation Study
  • Systematic Review

MeSH terms

  • Adenosine Triphosphate / analysis*
  • Bacteria / isolation & purification*
  • Bacteriological Techniques
  • Cholangiopancreatography, Endoscopic Retrograde / methods*
  • Duodenoscopes / microbiology*
  • Equipment Contamination*
  • Luminescent Measurements

Substances

  • Adenosine Triphosphate