Transcutaneous spinal direct current stimulation of the lumbar and sacral spinal cord: a modelling study

J Neural Eng. 2018 Jun;15(3):036008. doi: 10.1088/1741-2552/aaac38. Epub 2018 Feb 9.

Abstract

Objective: Our aim was to perform a computational study of the electric field (E-field) generated by transcutaneous spinal direct current stimulation (tsDCS) applied over the thoracic, lumbar and sacral spinal cord, in order to assess possible neuromodulatory effects on spinal cord circuitry related with lower limb functions.

Approach: A realistic volume conductor model of the human body consisting of 14 tissues was obtained from available databases. Rubber pad electrodes with a metallic connector and a conductive gel layer were modelled. The finite element (FE) method was used to calculate the E-field when a current of 2.5 mA was passed between two electrodes. The main characteristics of the E-field distributions in the spinal grey matter (spinal-GM) and spinal white matter (spinal-WM) were compared for seven montages, with the anode placed either over T10, T8 or L2 spinous processes (s.p.), and the cathode placed over right deltoid (rD), umbilicus (U) and right iliac crest (rIC) areas or T8 s.p. Anisotropic conductivity of spinal-WM and of a group of dorsal muscles near the vertebral column was considered.

Main results: The average E-field magnitude was predicted to be above 0.15 V m-1 in spinal cord regions located between the electrodes. L2-T8 and T8-rIC montages resulted in the highest E-field magnitudes in lumbar and sacral spinal segments (>0.30 V m-1). E-field longitudinal component is 3 to 6 times higher than the ventral-dorsal and right-left components in both the spinal-GM and WM. Anatomical features such as CSF narrowing due to vertebrae bony edges or disks intrusions in the spinal canal correlate with local maxima positions.

Significance: Computational modelling studies can provide detailed information regarding the electric field in the spinal cord during tsDCS. They are important to guide the design of clinical tsDCS protocols that optimize stimulation of application-specific spinal targets.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Electrodes, Implanted
  • Humans
  • Lumbar Vertebrae / physiology*
  • Male
  • Models, Neurological*
  • Sacrum / physiology*
  • Spinal Cord / physiology*
  • Transcranial Direct Current Stimulation / instrumentation
  • Transcranial Direct Current Stimulation / methods*
  • Transcutaneous Electric Nerve Stimulation / instrumentation
  • Transcutaneous Electric Nerve Stimulation / methods*