Experimental assessment of the Advanced Collapsed-cone Engine for scalp brachytherapy treatments

Brachytherapy. 2018 Mar-Apr;17(2):489-499. doi: 10.1016/j.brachy.2017.10.010. Epub 2017 Nov 24.

Abstract

Purpose: To experimentally assess the performance of the Advanced Collapsed-cone Engine (ACE) for 192Ir high-dose-rate brachytherapy treatment planning of nonmelanoma skin cancers of the scalp.

Methods and materials: A layered slab phantom was designed to model the head (skin, skull, and brain) and surface treatment mold using tissue equivalent materials. Six variations of the phantom were created by varying skin thickness, skull thickness, and size of air gap between the mold and skin. Treatment planning was initially performed using the Task Group 43 (TG-43) formalism with CT images of each phantom variation. Doses were recalculated using standard and high accuracy modes of ACE. The plans were delivered to Gafchromic EBT3 film placed between different layers of the phantom.

Results: Doses calculated by TG-43 and ACE and those measured by film agreed with each other at most locations within the phantoms. For a given phantom variation, average TG-43- and ACE-calculated doses were similar, with a maximum difference of (3 ± 12)% (k = 2). Compared to the film measurements, TG-43 and ACE overestimated the film-measured dose by (13 ± 12)% (k = 2) for one phantom variation below the skull layer.

Conclusions: TG-43- and ACE-calculated and film-measured doses were found to agree above the skull layer of the phantom, which is where the tumor would be located in a clinical case. ACE appears to underestimate the attenuation through bone relative to that measured by film; however, the dose to bone is below tolerance levels for this treatment.

Keywords: ACE; Dosimetry calculations; High-dose-rate (HDR); Scalp.

MeSH terms

  • Algorithms
  • Brachytherapy / methods*
  • Head and Neck Neoplasms / radiotherapy*
  • Humans
  • Iridium Radioisotopes
  • Monte Carlo Method
  • Phantoms, Imaging
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Scalp*
  • Skin Neoplasms / radiotherapy*

Substances

  • Iridium Radioisotopes
  • Iridium-192