Vector Flow Imaging Compared with Pulse Wave Doppler for Estimation of Peak Velocity in the Portal Vein

Ultrasound Med Biol. 2018 Mar;44(3):593-601. doi: 10.1016/j.ultrasmedbio.2017.10.015. Epub 2017 Dec 6.

Abstract

The study described here investigated whether angle-independent vector flow imaging (VFI) technique estimates peak velocities in the portal vein comparably to pulsed wave Doppler (PWD). Furthermore, intra- and inter-observer agreement was assessed in a substudy. VFI and PWD peak velocities were estimated with from intercostal and subcostal views for 32 healthy volunteers, and precision analyses were conducted. Blinded to estimates, three physicians rescanned 10 volunteers for intra- and inter-observer agreement analyses. The precision of VFI and PWD was 18% and 28% from an intercostal view and 23% and 77% from a subcostal view, respectively. Bias between VFI and PWD was 0.57 cm/s (p = 0.38) with an intercostal view and 9.89 cm/s (p <0.001) with a subcostal view. Intra- and inter-observer agreement was highest for VFI (inter-observer intra-class correlation coefficient: VFI 0.80, PWD 0.3; intra-observer intra-class correlation coefficient: VFI 0.90, PWD 0.69). Regardless of scan view, VFI was more precise than PWD.

Keywords: Abdominal ultrasound; Agreement analysis; Portal vein; Precision analysis; Pulsed wave Doppler; Vector flow imaging; Vector velocity.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Blood Flow Velocity / physiology
  • Female
  • Humans
  • Male
  • Middle Aged
  • Observer Variation
  • Portal Vein / diagnostic imaging
  • Portal Vein / physiology*
  • Reference Values
  • Reproducibility of Results
  • Ultrasonography / methods*
  • Ultrasonography, Doppler, Pulsed / methods