Common-path phase-sensitive optical coherence tomography provides enhanced phase stability and detection sensitivity for dynamic elastography

Biomed Opt Express. 2017 Oct 26;8(11):5253-5266. doi: 10.1364/BOE.8.005253. eCollection 2017 Nov 1.

Abstract

Phase-sensitive optical coherence elastography (PhS-OCE) is an emerging optical technique to quantify soft-tissue biomechanical properties. We implemented a common-path OCT design to enhance displacement sensitivity and optical phase stability for dynamic elastography imaging. The background phase stability was greater in common-path PhS-OCE (0.24 ± 0.07nm) than conventional PhS-OCE (1.60 ± 0.11μm). The coefficient of variation for surface displacement measurements using conventional PhS-OCE averaged 11% versus 2% for common-path PhS-OCE. Young's modulus estimates showed good precision (95% CIs) for tissue phantoms: 24.96 ± 2.18kPa (1% agar), 49.69 ± 4.87kPa (1.5% agar), and 116.08 ± 12.14kPa (2% agar), respectively. Common-path PhS-OCE effectively reduced the amplitude of background dynamic optical phase instability to a sub-nanometer level, which provided a larger dynamic detection range and higher detection sensitivity for surface displacement measurements than conventional PhS-OCE.

Keywords: (110.4500) Optical coherence tomography; (120.4570) Optical design of instruments; (120.5050) Phase measurement; (120.7280) Vibration analysis; Elastography; (170.3880) Medical and biological imaging.