The impact of a high-definition multileaf collimator for spine SBRT

J Appl Clin Med Phys. 2017 Nov;18(6):97-103. doi: 10.1002/acm2.12197. Epub 2017 Sep 27.

Abstract

Purpose: Advanced radiotherapy delivery systems designed for high-dose, high-precision treatments often come equipped with high-definition multi-leaf collimators (HD-MLC) aimed at more finely shaping radiation dose to the target. In this work, we study the effect of a high definition MLC on spine stereotactic body radiation therapy (SBRT) treatment plan quality and plan deliverability.

Methods and materials: Seventeen spine SBRT cases were planned with VMAT using a standard definition MLC (M120), HD-MLC, and HD-MLC with an added objective to reduce monitor units (MU). M120 plans were converted into plans deliverable on an HD-MLC using in-house software. Plan quality and plan deliverability as measured by portal dosimetry were compared among the three types of plans.

Results: Only minor differences were noted in plan quality between the M120 and HD-MLC plans. Plans generated with the HD-MLC tended to have better spinal cord sparing (3% reduction in maximum cord dose). HD-MLC plans on average had 12% more MU and 55% greater modulation complexity as defined by an in-house metric. HD-MLC plans also had significantly degraded deliverability. Of the VMAT arcs measured, 94% had lower gamma passing metrics when using the HD-MLC.

Conclusion: Modest improvements in plan quality were noted when switching from M120 to HD-MLC at the expense of significantly less accurate deliverability in some cases.

Keywords: HD-MLC; complexity; deliverability; spine SBRT.