Assessment of a photon-counting detector for a dual-energy C-arm angiographic system

Med Phys. 2017 Nov;44(11):5938-5948. doi: 10.1002/mp.12517. Epub 2017 Sep 22.

Abstract

Purpose: This article presents the implementation and assessment of photon-counting dual-energy x-ray detector technology for angiographic C-arm systems in interventional radiology.

Methods: A photon-counting detector was successfully integrated into a clinical C-arm CT system. Detector performance was assessed using image uniformity metrics in both 2D projections and 3D cone-beam computed tomography (CBCT) images. Uniform exposure fields were acquired to analyze projection images and scans of a homogeneous cylinder phantom were taken to analyze 3D reconstructions. Image uniformity was assessed over a broad range of imaging parameters.

Results: Detector calibration greatly improved image uniformity, reducing image variation from 8.8% to 0.5% in an ideal scenario, but image uniformity degraded when imaging parameters varied strongly from values set at calibration: the tube voltage, low-high energy threshhold, and tube current had the greatest impact. Material discrimination and dynamic angiography capabilities were successfully demonstrated in separate phantom and in vivo experiments.

Conclusion: The uniformity results identified major factors degrading image quality. The quantitative results will guide selection of calibration points to mitigate the loss of uniformity. The unique combination of dual-energy and fluoroscopy imaging capabilities with a flat-panel photon-counting detector may enable new applications in interventional radiology.

Keywords: C-arm; computed tomography; dual-energy; photon-counting; x-ray detectors.

MeSH terms

  • Angiography / instrumentation*
  • Calibration
  • Imaging, Three-Dimensional
  • Phantoms, Imaging
  • Photons*
  • Temperature
  • Tomography, X-Ray Computed