Peak Inspiratory Flow Rate in Chronic Obstructive Pulmonary Disease: Implications for Dry Powder Inhalers

J Aerosol Med Pulm Drug Deliv. 2017 Dec;30(6):381-387. doi: 10.1089/jamp.2017.1416. Epub 2017 Sep 21.

Abstract

Chronic obstructive pulmonary disease (COPD) is the third leading cause of death in the United States with a significant economic burden related to hospital admissions for exacerbations. One of the primary treatment modalities for COPD is medications delivered through breath-actuated dry powdered inhalers (DPIs). For users to successfully receive inhaled medication, they must inhale with enough flow to overcome the internal resistance of the device, leading to deaggregation of the medication powder. Peak inspiratory flow rate (PIFR) is the maximal flow rate obtained during an inspiratory maneuver. PIFR measurement can be impacted by the internal resistance of the device, which varies with device design. Many devices require a PIFR >60 L/min for adequate medication dispersal, while others appear to have adequate drug deaggregation with a PIFR >30 L/min. Studies have shown PIFRs are reduced among females and decrease with age, without a clear correlation between forced expiratory volume in 1 second and PIFR. PIFR can be reduced at the time of COPD exacerbation. Recent data suggest that reduced PIFR may be associated with worse COPD-related symptom burden, increased odds of COPD-related hospital readmissions, and improved responsiveness to nebulized therapy. This review article aims to examine the physiology and clinical correlations of PIFR, as well as review published studies related to PIFR with DPIs used to treat COPD.

Keywords: COPD; dry powder inhaler; peak inspiratory flow.

Publication types

  • Review

MeSH terms

  • Administration, Inhalation
  • Drug Delivery Systems*
  • Dry Powder Inhalers*
  • Equipment Design
  • Female
  • Forced Expiratory Volume
  • Humans
  • Inhalation / physiology
  • Male
  • Pulmonary Disease, Chronic Obstructive / drug therapy*