Longitudinal dialysis adequacy and clinical performance of the VectorFlow hemodialysis catheter: a prospective study

J Vasc Access. 2017 Nov 17;18(6):492-497. doi: 10.5301/jva.5000784. Epub 2017 Aug 26.

Abstract

Purpose: To report clinical performance and longitudinal assessment of hemodialysis adequacy with the Arrow-Clark VectorFlow catheter, a symmetrical-tip device with a distal lumen configuration designed to reduce platelet shear stress and catheter thrombosis.

Methods and materials: We prospectively enrolled patients who required de novo placement of a chronic tunneled catheter for hemodialysis or exchange of a dysfunctional catheter as part of an Institutional Review Board (IRB)-approved protocol. Catheter patency, Kt/V, mean blood-flow (Qb), and pump pressures were obtained at baseline and at monthly intervals to 90 days.

Results: Forty-six subjects were enrolled into the study. During the 90-day observation period, maximum blood-flow rate averaged 355-398 mL/minute; mean Qb averaged 333-392 mL/minute. Mean Kt/V values were consistently ≥1.5. Dwell-time was 15-114 days, for a total of 2997 catheter days (mean 71.4 days). Excluding patients who died during the study and those receiving surgical access, overall intervention-free catheter patency rate was 94.9%, 92.2% and 88.8% at days 30, 60, 90, respectively. There were no acute complications. During the follow-up period, three patients developed complications (6.5%). Two catheter infections occurred (0.7/1000 catheter days) and one catheter malfunctioned; a rate of 1.0/ 1000 catheter days for all complications.

Conclusions: The VectorFlow catheter produced safe, effective hemodialysis with Kt/V ≥1.5. A single catheter occlusion occurred and a low rate of infection was seen. Results support the hypothesis that the VectorFlow design reduces thrombogenic risk during clinical performance.

Publication types

  • Observational Study

MeSH terms

  • Aged
  • Blood Flow Velocity
  • Catheter Obstruction / etiology
  • Catheter-Related Infections / etiology
  • Catheterization, Central Venous / adverse effects
  • Catheterization, Central Venous / instrumentation*
  • Catheters, Indwelling*
  • Central Venous Catheters*
  • Equipment Design
  • Equipment Failure
  • Female
  • Humans
  • Kidney Failure, Chronic / diagnosis
  • Kidney Failure, Chronic / physiopathology
  • Kidney Failure, Chronic / therapy*
  • Male
  • Middle Aged
  • Prospective Studies
  • Regional Blood Flow
  • Renal Dialysis / adverse effects
  • Renal Dialysis / instrumentation*
  • Risk Factors
  • Texas
  • Time Factors
  • Treatment Outcome