DMF-MALDI: droplet based microfluidic combined to MALDI-TOF for focused peptide detection

Sci Rep. 2017 Jul 28;7(1):6756. doi: 10.1038/s41598-017-06660-8.

Abstract

We present an automated droplet microfluidic system (DMF) to generate monitored nanoliter aqueous droplets in oil and their deposition on a commercial stainless steel plate for MALDI-TOF analysis of peptides or protein digests. We demonstrate that DMF-MALDI combination focuses the analyte on the MALDI plate, increasing considerably the homogeneity of the dried material. This results in a 30times enhanced MALDI-TOF MS signal for a model peptide, allowing a significant improvement of the detection sensitivity limit (down to few tens of attomoles). Moreover, positive detection can be achieved from sub-nanomolar peptides solutions and better overall protein sequence coverages are obtained from few tens attomoles of protein digest. These results make DMF-MALDI a promising approach for the treatment of peptides samples as well as a key component for an integrated approach in the proteomic field.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Angiotensin II / analysis
  • Animals
  • Humans
  • Microfluidics*
  • Peptides / analysis*
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization / methods*
  • Sus scrofa

Substances

  • Peptides
  • Angiotensin II