Simple algorithm for the correction of MRI image artefacts due to random phase fluctuations

Magn Reson Imaging. 2017 Dec:44:55-59. doi: 10.1016/j.mri.2017.07.023. Epub 2017 Jul 24.

Abstract

Purpose: Fast Field-Cycling (FFC) MRI is a novel technology that allows varying the main magnetic field B0 during the pulse sequence, from the nominal field (usually hundreds of millitesla) down to Earth's field or below. This technique uses resistive magnets powered by fast amplifiers. One of the challenges with this method is to stabilise the magnetic field during the acquisition of the NMR signal. Indeed, a typical consequence of field instability is small, random phase variations between each line of k-space resulting in artefacts, similar to those which occur due to homogeneous motion but harder to correct as no assumption can be made about the phase error, which appears completely random. Here we propose an algorithm that can correct for the random phase variations induced by field instabilities without prior knowledge about the phase error.

Methods: The algorithm exploits the fact that ghosts caused by field instability manifest in image regions which should be signal free. The algorithm minimises the signal in the background by finding an optimum phase correction for each line of k-space and repeats the operation until the result converges, leaving the background free of signal.

Conclusion: We showed the conditions for which the algorithm is robust and successfully applied it on images acquired on FFC-MRI scanners. The same algorithm can be used for various applications other than Fast Field-Cycling MRI.

Keywords: Correction algorithm; Fast field-cycling MRI; Phase encode artefact; Post-processing.

MeSH terms

  • Algorithms
  • Artifacts
  • Computer Simulation
  • Image Processing, Computer-Assisted*
  • Magnetic Resonance Imaging*
  • Models, Statistical
  • Motion*
  • Normal Distribution
  • Phantoms, Imaging
  • Signal Processing, Computer-Assisted