Failure analysis of eleven Gates Glidden drills that fractured intraorally during post space preparation. A retrieval analysis study

Biomed Tech (Berl). 2018 Jul 26;63(4):407-412. doi: 10.1515/bmt-2016-0245.

Abstract

The purpose of this study was to determine the failure mechanism of clinically failed Gates Glidden (GG) drills. Eleven retrieved GG drills (sizes #1 to #3) which fractured during root canal preparation were collected and the fracture location was recorded based on macroscopic observation. All fracture surfaces were investigated by a SEM. Then the fractured parts were embedded in acrylic resin and after metallographic preparation, the microstructure and elemental composition was evaluated by SEM and EDS. The Vickers hardness (HV) of all specimens was also determined. Macroscopic examination and SEM analysis showed that the drills failed near the hand piece end by torsional fatigue with fatigue cracks initiating at several locations around the circumference and propagating toward the center. Final fracture followed by a tensile overloading at the central region of cross section. Microstructural analysis, hardness measurements and EDS show that the drills are made of a martensitic stainless steel like AISI 440C. Based on the findings of this study, clinicians should expect fatigue fracture of GG drills that have small size during root canal preparation. Selection of a more fatigue resistant stainless steel alloy and enhancing the instrument design might reduce the incidence of quasi-cleavage fracture on GG drills.

Keywords: Gates Glidden; SEM/EDS; failure analysis; fractography; torsional fatigue.

MeSH terms

  • Alloys*
  • Hardness
  • Stainless Steel / standards*

Substances

  • Alloys
  • Stainless Steel