Analysis of polychlorodibenzo-p-dioxins and polychlorodibenzofurans in stationary source emissions in GC-MS/MS using hydrogen as the carrier gas

J Sep Sci. 2017 Sep;40(17):3469-3478. doi: 10.1002/jssc.201700026. Epub 2017 Aug 14.

Abstract

The European and American methods for the determination of polychlorodibenzo-p-dioxins and polychlorodibenzofurans in stationary source emissions require a high-resolution mass spectrometer detector. At the same time, all of the 17 toxic 2,3,7,8-chlorine-substituted isomers have to be separated by an appropriate chromatographic method. The resulting analysis has long runtimes and sometimes a double-column run is needed, which results in a huge waste of time, money and manual labor. To obtain a fast and cost-effective but still reliable analytical system, a single-column gas chromatography with hydrogen as carrier gas and tandem mass spectrometry detection is proposed. The use of hydrogen as the carrier gas is a good choice to shorten runtimes and improve the chromatographic resolution, while tandem mass spectrometry is a well-known alternative for this analysis. All the chromatographic and mass spectrometric parameters fill the requirements of the reference methods with a 35% reduction of runtimes. The accuracy is tested with three fly ash samples of a proficiency interlaboratory test. A good correlation between the results is obtained (R2 = 0.992, slope = 0.9675), and no coelutions are noted. The system robustness is tested during 5 years of constant use and the maximal relative standard deviation of the relative response factor is 18.8%.

Keywords: dioxins; emissions; furans; hydrogen; tandem mass spectrometry.