Development of an instrumented spinal cord surrogate using optical fibers: A feasibility study

Med Eng Phys. 2017 Oct:48:212-216. doi: 10.1016/j.medengphy.2017.06.033. Epub 2017 Jul 4.

Abstract

In vitro replication of traumatic spinal cord injury is necessary to understand its biomechanics and to improve animal models. During a traumatic spinal cord injury, the spinal cord withstands an impaction at high velocity. In order to fully assess the impaction, the use of spinal canal occlusion sensor is necessary. A physical spinal cord surrogate is also often used to simulate the presence of the spinal cord and its surrounding structures. In this study, an instrumented physical spinal cord surrogate is presented and validated. The sensing is based on light transmission loss observed in embedded bare optical fibers subjected to bending. The instrumented surrogate exhibits similar mechanical properties under static compression compared to fresh porcine spinal cords. The instrumented surrogate has a compression sensing threshold of 40% that matches the smallest compression values leading to neurological injuries. The signal obtained from the sensor allows calculating the compression of the spinal cord surrogate with a maximum of 5% deviation. Excellent repeatability was also observed under repetitive loading. The proposed instrumented spinal cord surrogate is promising with satisfying mechanical properties and good sensing capability. It is the first attempt at proposing a method to assess the internal loads sustained by the spinal cord during a traumatic injury.

Keywords: Instrumented surrogate; Optical fiber sensing; Physical spinal cord surrogate; Traumatic spinal cord injury.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Compressive Strength
  • Feasibility Studies
  • Models, Anatomic
  • Optical Fibers*
  • Printing, Three-Dimensional
  • Spinal Cord*