On-line coupling of molecularly imprinted solid phase extraction with liquid chromatography for the fast determination of coumarins from complex samples

J Pharm Biomed Anal. 2017 Oct 25:145:144-150. doi: 10.1016/j.jpba.2017.06.033. Epub 2017 Jun 20.

Abstract

In this work, an on-line SPE-HPLC method with spectrophotometric detection was developed for the determination of coumarins in complex samples. For the on-line cleanup of samples, a molecularly imprinted polymer was packed into the column cartridge and coupled directly with HPLC (MISPE-HPLC) using a column switching system. The separation of coumarins was performed on a C18 core-shell column (100×4.6mm, 5μm) with a mobile phase consisting of 0.3% acetic acid/acetonitrile with gradient elution at a flow-rate of 1mLmin-1. The total time of the whole analytical run, including the extraction step, was 13.25min. The on-line MISPE-HPLC method was optimized and validated. The results showed good linearity (0.10-100μgmL-1) with correlation coefficients higher than 0.99. The LOD values were from 0.03 to 0.15μgmL-1. The proposed method was successfully applied for analysis of real samples (Cassia cinnamon, chamomile tea, and Tokaj specialty wines) and obtained recoveries varied from 78.7% to 112.2% with an RSD less than 9%.

Keywords: Chromatography; Column switching system; Coumarins; Molecularly imprinted polymers; On-line solid phase extraction.

MeSH terms

  • Chromatography, High Pressure Liquid
  • Coumarins
  • Molecular Imprinting
  • Online Systems
  • Polymers
  • Solid Phase Extraction*

Substances

  • Coumarins
  • Polymers