Volume guaranteed? Accuracy of a volume-targeted ventilation mode in infants

Arch Dis Child Fetal Neonatal Ed. 2018 Mar;103(2):F120-F125. doi: 10.1136/archdischild-2017-312640. Epub 2017 Jun 28.

Abstract

Objectives: Volume-targeted ventilation (VTV) is widely used and may reduce lung injury, but this assumes the clinically set tidal volume (VTset) is accurately delivered. This prospective observational study aimed to determine the relationship between VTset, expiratory VT (VTe) and endotracheal tube leak in a modern neonatal -volume-targeted ventilator (VTV) and the resultant partial arterial pressure of carbon dioxide (PaCO2) relationship with and without VTV.

Design: Continuous inflations were recorded for 24 hours in 100 infants, mean (SD) 34 (4) weeks gestation and 2483 (985) g birth weight, receiving synchronised mechanical ventilation (SLE5000, SLE, UK) with or without VTV and either the manufacturer's V4 (n=50) or newer V5 (n=50) VTV algorithm. The VTset, VTe and leak were determined for each inflation (maximum 90 000/infant). If PaCO2 was sampled (maximum of 2 per infant), this was compared with the average VTe data from the preceding 15 min.

Results: A total of 7 497 137 inflations were analysed. With VTV enabled (77 infants), the VTset-VTe bias (95% CI) was 0.03 (-0.12 to 0.19) mL/kg, with a median of 80% of VTe being ±1.0 mL/kg of VTset. Endotracheal tube leak up to 30% influenced VTset-VTe bias with the V4 (r2=-0.64, p<0.0001; linear regression) but not V5 algorithm (r2=0.04, p=0.21). There was an inverse linear relationship between VTe and PaCO2 without VTV (r2=0.26, p=0.004), but not with VTV (r2=0.04, p=0.10), and less PaCO2 within 40-60 mm Hg, 53% versus 72%, relative risk (95% CI) 1.7 (1.0 to 2.9).

Conclusion: VTV was accurate and reliable even with moderate leak and PaCO2 more stable. VTV algorithm differences may exist in other devices.

Keywords: infant; lung injury; mechanical ventilation; tidal volume; volume targeted ventilation.

Publication types

  • Observational Study

MeSH terms

  • Algorithms
  • Carbon Dioxide / blood
  • Female
  • Gestational Age
  • Humans
  • Infant, Newborn
  • Infant, Premature*
  • Intensive Care Units, Neonatal
  • Intubation, Intratracheal
  • Male
  • Prospective Studies
  • Respiration, Artificial / instrumentation*
  • Respiration, Artificial / methods*
  • Respiration, Artificial / standards
  • Tidal Volume*

Substances

  • Carbon Dioxide