On stress/strain shielding and the material stiffness paradigm for dental implants

Clin Implant Dent Relat Res. 2017 Oct;19(5):935-943. doi: 10.1111/cid.12509. Epub 2017 Jun 13.

Abstract

Background: Stress shielding considerations suggest that the dental implant material's compliance should be matched to that of the host bone. However, this belief has not been confirmed from a general perspective, either clinically or numerically.

Purpose: To characterize the influence of the implant stiffness on its functionality using the failure envelope concept that examines all possible combinations of mechanical load and application angle for selected stress, strain and displacement-based bone failure criteria. Those criteria represent bone yielding, remodeling, and implant primary stability, respectively MATERIALS AND METHODS: We performed numerical simulations to generate failure envelopes for all possible loading configurations of dental implants, with stiffness ranging from very low (polymer) to extremely high, through that of bone, titanium, and ceramics.

Results: Irrespective of the failure criterion, stiffer implants allow for improved implant functionality. The latter reduces with increasing compliance, while the trabecular bone experiences higher strains, albeit of an overall small level. Micromotions remain quite small irrespective of the implant's stiffness.

Conclusion: The current paradigm favoring reduced implant material's stiffness out of concern for stress or strain shielding, or even excessive micromotions, is not supported by the present calculations, that point exactly to the opposite.

Keywords: Young's modulus; failure envelope; finite element; implant material; micromotions; stress/strain shielding.

MeSH terms

  • Dental Implants*
  • Dental Materials*
  • Dental Stress Analysis*
  • Materials Testing*
  • Stress, Mechanical*