Long-term outcome of bioresorbable vascular scaffolds for the treatment of coronary artery disease: a meta-analysis of RCTs

BMC Cardiovasc Disord. 2017 Jun 7;17(1):147. doi: 10.1186/s12872-017-0586-2.

Abstract

Background: Coronary bioresorbable scaffolds (BRS) were developed to overcome the limitations of standard metallic stents, especially to address late events after percutaneous coronary interventions. The aim of this meta-analysis was to evaluate the efficacy and safety of BRS, compared with Everolimus-eluting stents (EES), using the data available from randomized trials, with a focus on long-term outcomes.

Methods: Published randomized trials comparing BRS to EES for the treatment of coronary artery disease were searched for within PubMed, Cochrane Library and Scopus electronic databases up to April 4th 2017. The summary measure used was odds ratio (OR) with 95% confidence intervals.

Results: A total of 5 studies were eligible, including 5219 patients. At 2 years, BRS was associated with higher rates of target lesion failure (9.4% vs 7.2%; OR = 1.33; 95% CI 1.07 to 1.63; p = 0.008) and device thrombosis (2.3% vs 0.7%; OR = 3.22; 95% CI 1.86 to 5.57; p < 0.0001) compared with EES. The incidence of both early (within 30 days after implantation, 1.1% vs 0.5%, OR 1.97, 95% CI 1.02 to 3.81; p = 0.05) and very-late device thrombosis (>1 year, 0.6% vs 0.1%, OR 4.03, 95% CI 1.37 to 11.82; p = 0.01) was higher with BRS compared with EES.

Conclusions: BRS may be associated with worse two-years clinical outcomes compared with EES in patients with coronary artery disease.

Keywords: Bioresorbable vascular scaffold; Stent thrombosis; Target lesion failure.

Publication types

  • Meta-Analysis
  • Review

MeSH terms

  • Absorbable Implants*
  • Cardiovascular Agents / administration & dosage
  • Chi-Square Distribution
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / therapy*
  • Coronary Thrombosis / etiology
  • Drug-Eluting Stents*
  • Everolimus / administration & dosage
  • Humans
  • Odds Ratio
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Prosthesis Design
  • Randomized Controlled Trials as Topic
  • Risk Factors
  • Time Factors
  • Treatment Outcome

Substances

  • Cardiovascular Agents
  • Everolimus