Quantitative pulsed CEST-MRI at a clinical 3T MRI system

MAGMA. 2017 Oct;30(5):505-516. doi: 10.1007/s10334-017-0625-0. Epub 2017 May 31.

Abstract

Objectives: The goal of this study was to quantify CEST related parameters such as chemical exchange rate and fractional concentration of exchanging protons at a clinical 3T scanner. For this purpose, two CEST quantification approaches-the AREX metric (for 'apparent exchange dependent relaxation'), and the AREX-based Ω-plot method were used. In addition, two different pulsed RF irradiation schemes, using Gaussian-shaped and spin-lock pulses, were compared.

Materials and methods: Numerical simulations as well as MRI measurements in phantoms were performed. For simulations, the Bloch-McConnell equations were solved using a two-pool exchange model. MR experiments were performed on a clinical 3T MRI scanner using a cylindrical phantom filled with creatine solution at different pH values and different concentrations.

Results: The validity of the Ω-plot method and the AREX approach using spin-lock preparation for determination of the quantitative CEST parameters was demonstrated. Especially promising results were achieved for the Ω-plot method when the spin-lock preparation was employed.

Conclusion: Pulsed CEST at 3T could be used to quantify parameters such as exchange rate constants and concentrations of protons exchanging with free water. In the future this technique might be used to estimate the exchange rates and concentrations of biochemical substances in human tissues in vivo.

Keywords: Chemical exchange saturation transfer (CEST); Creatine; Magnetic resonance imaging (MRI).

Publication types

  • Evaluation Study

MeSH terms

  • Computer Simulation
  • Humans
  • Hydrogen-Ion Concentration
  • Magnetic Fields
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Imaging / statistics & numerical data
  • Phantoms, Imaging
  • Protons
  • Radio Waves
  • Water

Substances

  • Protons
  • Water