Modeling the detection efficiency of an HP-Ge detector for use in boron neutron capture therapy

Appl Radiat Isot. 2017 Jul:125:80-85. doi: 10.1016/j.apradiso.2017.03.027. Epub 2017 Apr 2.

Abstract

The multi-foil method is commonly used to determine upon an energy spectrum of neutrons in boron neutron capture therapy. The method requires to measure the radioactivation of the foils. This study develops a simple modeling procedure of a high-purity Ge detector, which is used to measure the radioactivation, in order to calculate the detection efficiency with GEANT4. By changing four parameters from their manufacturing specifications of the detector, the simulated detection efficiency is able to reproduce the actual detection efficiency.

Keywords: BNCT; Detection efficiency; HP-Ge detector; Modeling procedure; Multi-foil method.