A Delayed-Excitation Data Acquisition Method for High-Frequency Ultrasound Imaging

IEEE Trans Biomed Eng. 2018 Jan;65(1):15-20. doi: 10.1109/TBME.2017.2687948. Epub 2017 Mar 28.

Abstract

High-frequency ultrasound imaging (at >20 MHz) has gained widespread attention due to its high spatial resolution being useful for basic cardiovascular and cancer research involving small animals. The sampling rate of the analog-to-digital converter in a high-frequency ultrasound system usually needs to be higher than 120 MHz in order to satisfy the Nyquist sampling-rate requirement. However, the sampling rate is typically within the range of 40-60 MHz in a traditional ultrasound system, and so we propose a delayed-excitation method for performing high-frequency ultrasound imaging with a traditional data acquisition scheme. In this method, the transmitted pulse is delayed by a certain time period so that the ultrasound echo data are aligned into high-sampling-rate slots. Wire and tissue-mimicking phantoms were imaged to evaluate the performance of the proposed method, whereas a porcine small-intestine specimen and an excised rabbit eyeball were used for in vitro imaging evaluations. The test results demonstrate that the proposed method allows high-frequency ultrasound imaging to be implemented using a traditional ultrasound sampling system.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Eye / diagnostic imaging
  • Image Processing, Computer-Assisted / methods*
  • Intestine, Small / diagnostic imaging
  • Phantoms, Imaging
  • Rabbits
  • Signal Processing, Computer-Assisted*
  • Swine
  • Ultrasonography / methods*