A column switching ultrahigh-performance liquid chromatography-tandem mass spectrometry method to determine anandamide and 2-arachidonoylglycerol in plasma samples

Anal Bioanal Chem. 2017 May;409(14):3587-3596. doi: 10.1007/s00216-017-0300-3. Epub 2017 Mar 23.

Abstract

This study reports a fast, sensitive, and selective column switching ultrahigh-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method to determine the endocannabinoids (eCBs), anandamide (AEA), and 2-arachidonoylglycerol (2-AG) in plasma samples. This bidimensional system used a restricted access media column (RP-8 ADS, 25 mm × 4 mm × 25 μM) in the first dimension and a core-shell Kinetex C18 (100 mm × 2, 1.7 mm × 1 μM) column in the second dimension, followed by detection in a mass spectrometer triple quadrupole (multiple reactions monitoring mode) operating in the positive mode. RP-8 ADS was used for trace enrichment of eCBs (reverse phase partitioning) and macromolecular matrix size exclusion; the core-shell column was used for the chromatographic separation. The column switching UHPLC-MS/MS method presented a linear range spanning from 0.1 ng mL-1 (LOQ) to 6 ng mL-1 for AEA and from 0.04 ng mL-1 (LOQ) to 10 ng mL-1 for 2-AG. Excluding the LLOQ values, the precision assays provided coefficients of variation lower than 8% and accuracy with relative standard error values lower than 14%. Neither carryover nor matrix effects were detected. This high-throughput column switching method compared to conventional methods is time saving as it involves fewer steps, consumes less solvent, and presents lower LLOQ. The column switching UHPLC-MS/MS method was successfully applied to determine AEA and 2-AG in plasma samples obtained from Alzheimer's disease patients. Graphical abstract A column switching ultra high-performance liquid chromatography-tandem mass spectrometry method using RP-8 ADS column and core shell column to determine endocannabinoids in plasma samples.

Keywords: Column switching; Endocannabinoids; Liquid chromatography-tandem mass spectrometry; Plasma samples.

Publication types

  • Validation Study

MeSH terms

  • Alzheimer Disease / blood
  • Arachidonic Acids / blood*
  • Chromatography, High Pressure Liquid / instrumentation
  • Chromatography, High Pressure Liquid / methods*
  • Endocannabinoids / blood*
  • Equipment Design
  • Glycerides / blood*
  • Humans
  • Limit of Detection
  • Polyunsaturated Alkamides / blood*
  • Reproducibility of Results
  • Tandem Mass Spectrometry / instrumentation
  • Tandem Mass Spectrometry / methods*

Substances

  • Arachidonic Acids
  • Endocannabinoids
  • Glycerides
  • Polyunsaturated Alkamides
  • glyceryl 2-arachidonate
  • anandamide