Effects of coplanar shielding for high field MRI

Annu Int Conf IEEE Eng Med Biol Soc. 2016 Aug:2016:6250-6253. doi: 10.1109/EMBC.2016.7592157.

Abstract

This work investigates the efficacy of "coplanar shielding," in which copper shields are oriented concentric and coplanar to the RF coils rather than implemented as a full ground plane behind them. Following FDTD simulations to determine optimal shielding parameters, two coil geometries were constructed: a circular loop surface coil and a half-volume five-element receive array. Each was evaluated using bench measurements with and without coplanar shielding. Imaging, including accelerated SENSE imaging, was performed with the shielded and unshielded receive arrays on a whole-body 7T scanner. Results from modeled and fabricated coils showed good agreement with improvements in Q factors for all cases. Imaging showed substantial improvements in SNR and g-factors for the coplanar shielded array.

MeSH terms

  • Magnetic Resonance Imaging* / instrumentation
  • Magnetic Resonance Imaging* / methods
  • Models, Theoretical
  • Phantoms, Imaging