Accuracy of computer-aided diagnosis based on narrow-band imaging endocytoscopy for diagnosing colorectal lesions: comparison with experts

Int J Comput Assist Radiol Surg. 2017 May;12(5):757-766. doi: 10.1007/s11548-017-1542-4. Epub 2017 Feb 28.

Abstract

Purpose: Real-time characterization of colorectal lesions during colonoscopy is important for reducing medical costs, given that the need for a pathological diagnosis can be omitted if the accuracy of the diagnostic modality is sufficiently high. However, it is sometimes difficult for community-based gastroenterologists to achieve the required level of diagnostic accuracy. In this regard, we developed a computer-aided diagnosis (CAD) system based on endocytoscopy (EC) to evaluate cellular, glandular, and vessel structure atypia in vivo. The purpose of this study was to compare the diagnostic ability and efficacy of this CAD system with the performances of human expert and trainee endoscopists.

Methods: We developed a CAD system based on EC with narrow-band imaging that allowed microvascular evaluation without dye (ECV-CAD). The CAD algorithm was programmed based on texture analysis and provided a two-class diagnosis of neoplastic or non-neoplastic, with probabilities. We validated the diagnostic ability of the ECV-CAD system using 173 randomly selected EC images (49 non-neoplasms, 124 neoplasms). The images were evaluated by the CAD and by four expert endoscopists and three trainees. The diagnostic accuracies for distinguishing between neoplasms and non-neoplasms were calculated.

Results: ECV-CAD had higher overall diagnostic accuracy than trainees (87.8 vs 63.4%; [Formula: see text]), but similar to experts (87.8 vs 84.2%; [Formula: see text]). With regard to high-confidence cases, the overall accuracy of ECV-CAD was also higher than trainees (93.5 vs 71.7%; [Formula: see text]) and comparable to experts (93.5 vs 90.8%; [Formula: see text]).

Conclusions: ECV-CAD showed better diagnostic accuracy than trainee endoscopists and was comparable to that of experts. ECV-CAD could thus be a powerful decision-making tool for less-experienced endoscopists.

Keywords: Colon polyp; Colonoscopy; Computer-aided diagnosis; Endocytoscopy; Magnifying endoscopy; Narrow-band imaging.

MeSH terms

  • Aged
  • Algorithms
  • Colonic Polyps / diagnostic imaging*
  • Colonoscopy / methods*
  • Colorectal Neoplasms / diagnostic imaging*
  • Diagnosis, Computer-Assisted / methods*
  • Endoscopy / methods*
  • Female
  • Humans
  • Image Processing, Computer-Assisted
  • Male
  • Middle Aged
  • Narrow Band Imaging
  • Pattern Recognition, Automated
  • Probability
  • Reproducibility of Results
  • Retrospective Studies
  • Software
  • Treatment Outcome