Patency of Saphenous Vein Grafts Using the PAS-Port System During Coronary Artery Bypass Surgery

Ann Thorac Surg. 2017 Aug;104(2):560-567. doi: 10.1016/j.athoracsur.2016.11.014. Epub 2017 Feb 21.

Abstract

Background: Several proximal anastomosis devices have been developed to shorten the time required for a proximal anastomosis and to avoid aortic cross-/side-clamping during coronary artery bypass grafting. This study retrospectively examined the patency of saphenous vein grafts (SVGs) using the PAS-Port System (Cardia Inc, Redwood City, CA).

Methods: From 2004 to 2014, 451 patients underwent coronary artery bypass graft operations requiring at least 1 proximal anastomosis using a PAS-Port device. A total of 802 PAS-Port devices were used, and 95.0% (762 of 802) were implanted successfully. Among the successfully implanted anastomoses, 76.8% (585 of 762) were evaluated using coronary angiography or multidimensional computed tomography, or both. The evaluations were performed between postoperative days 4 and 3,182 (mean, 319 ± 624 days). The early (1 to 365 days) and the midterm to long-term (more than 366 days) occlusion rates were examined. A complete postoperative clinical course was recorded for 70.7% of the patients.

Results: Overall, 93.8% (549 of 585) of the device-dependent SVGs were patent. The patency rates of device-dependent SVGs that were 1, 2, 3, 4, 5, 6, 7, and 8 years old were 90.1% ± 1.8%, 87.1% ± 2.3%, 86.1% ± 2.5%, 82.9% ± 3.3%, 80.6% ± 3.9%, 77.2% ± 5.0%, 77.2% ± 5.0%, and 70.2% ± 8.1%, respectively. The longest follow-up period was 3,182 days (8.7 years). The occlusion rate for device-dependent SVGs tended to decrease as the number of patients accumulated.

Conclusions: The PAS-Port system provided acceptable SVG patency and clinical outcome for the early and midterm to long-term. There may be a learning curve for the use of PAS-Port device that affects the device-dependent SVG patency.

MeSH terms

  • Aged
  • Coronary Angiography
  • Coronary Artery Bypass / methods*
  • Coronary Artery Disease / diagnosis
  • Coronary Artery Disease / mortality
  • Coronary Artery Disease / surgery*
  • Equipment Design
  • Female
  • Follow-Up Studies
  • Graft Occlusion, Vascular / diagnosis
  • Graft Occlusion, Vascular / epidemiology
  • Graft Occlusion, Vascular / physiopathology*
  • Humans
  • Incidence
  • Japan / epidemiology
  • Male
  • Monitoring, Intraoperative / instrumentation*
  • Multidetector Computed Tomography
  • Retrospective Studies
  • Risk Factors
  • Saphenous Vein / diagnostic imaging
  • Saphenous Vein / physiopathology*
  • Saphenous Vein / transplantation
  • Survival Rate / trends
  • Vascular Patency*