Comparison of visual outcomes after bilateral implantation of extended range of vision and trifocal intraocular lenses

Eur J Ophthalmol. 2017 Jun 26;27(4):460-465. doi: 10.5301/ejo.5000935. Epub 2017 Jan 28.

Abstract

Purpose: To compare visual outcomes after cataract surgery with bilateral implantation of 2 intraocular lenses (IOLs): extended range of vision and trifocal.

Methods: Each group of this prospective study comprised 40 eyes (20 patients). Phacoemulsification followed by bilateral implantation of a FineVision IOL (group 1) or a Symfony IOL (group 2) was performed. The following outcomes were assessed up to 1 year postoperatively: binocular uncorrected distance visual acuity (UDVA), binocular uncorrected intermediate visual acuity (UIVA) at 60 cm, binocular uncorrected near visual acuity (UNVA) at 40 cm, spherical equivalent (SE) refraction, defocus curves, mesopic and photopic contrast sensitivity, halometry, posterior capsule opacification (PCO), and responses to a patient questionnaire.

Results: The mean binocular values in group 1 and group 2, respectively, were SE -0.15 ± 0.25 D and -0.19 ± 0.18 D; UDVA 0.01 ± 0.03 logMAR and 0.01 ± 0.02 logMAR; UIVA 0.11 ± 0.08 logMAR and 0.09 ± 0.08 logMAR; UNVA 0.06 ± 0.07 logMAR and 0.17 ± 0.06 logMAR. Difference in UNVA between IOLs (p<0.05) was statistically significant. There were no significant differences in contrast sensitivity, halometry, or PCO between groups. Defocus curves were similar between groups from 0 D to -2 D, but showed significant differences from -2.50 D to -4.00 D (p<0.05).

Conclusions: Both IOLs provided excellent distance and intermediate visual outcomes. The FineVision IOL showed better near visual acuity. Predictability of the refractive results and optical performance were excellent; all patients achieved spectacle independence. The 2 IOLs gave similar and good contrast sensitivity in photopic and mesopic conditions and low perception of halos by patients.

MeSH terms

  • Aged
  • Capsule Opacification / pathology
  • Color Vision
  • Contrast Sensitivity / physiology
  • Eyeglasses
  • Female
  • Humans
  • Lens Implantation, Intraocular*
  • Lens, Crystalline / physiopathology
  • Lenses, Intraocular*
  • Male
  • Middle Aged
  • Patient Satisfaction
  • Phacoemulsification*
  • Prospective Studies
  • Prosthesis Design
  • Refraction, Ocular / physiology
  • Surveys and Questionnaires
  • Vision Disorders
  • Vision, Binocular / physiology
  • Visual Acuity / physiology*