Optimizing disk registration algorithms for nanobeam electron diffraction strain mapping

Ultramicroscopy. 2017 May:176:170-176. doi: 10.1016/j.ultramic.2016.12.021. Epub 2017 Jan 28.

Abstract

Scanning nanobeam electron diffraction strain mapping is a technique by which the positions of diffracted disks sampled at the nanoscale over a crystalline sample can be used to reconstruct a strain map over a large area. However, it is important that the disk positions are measured accurately, as their positions relative to a reference are directly used to calculate strain. In this study, we compare several correlation methods using both simulated and experimental data in order to directly probe susceptibility to measurement error due to non-uniform diffracted disk illumination structure. We found that prefiltering the diffraction patterns with a Sobel filter before performing cross correlation or performing a square-root magnitude weighted phase correlation returned the best results when inner disk structure was present. We have tested these methods both on simulated datasets, and experimental data from unstrained silicon as well as a twin grain boundary in 304 stainless steel.

Keywords: Nanobeam electron diffraction; Strain measurement.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.