Reliability and validity of non-invasive determined haemoglobin mass and blood volumes

Clin Physiol Funct Imaging. 2018 Mar;38(2):240-245. doi: 10.1111/cpf.12406. Epub 2017 Jan 30.

Abstract

Introduction: The carbon monoxide (CO) rebreathing method used for the determination of haemoglobin mass (Hbmass ) is associated with blood sample analysis (in this study: Radiometer ABL800). As an alternative hereto the aim of the present study was to evaluate the use of a portable and non-invasive CO pulse oximeter (Rad-57).

Method: With simultaneous determination of CO in the circulation by ABL800 (%HbCO) and Rad-57 (SpCO), Hbmass and blood volume (BV) were determined in duplicates in 24 volunteers. Percentage of typical errors (%TE) within methods and linear correlations between the two procedures were computed.

Results: Hbmass (Rad-57 = 798 ± 230 g; ABL800 = 781 ± 192 g) and BV (Rad-57 = 5700 ± 1373 ml; ABL800 = 5581 ± 1096 ml) were similar between methods. However, the %TE for Hbmass was higher (P<0·001) for Rad-57 (5·84 ± 5·29%) than for ABL800 (1·35 ± 1·13%). Similarly, the %TE for BV was higher (P<0·001) for Rad-57 (6·06 ± 5·76%) than for ABL800 (1·48 ± 1·25%). Lower (P<0·05) correlation coefficients between the methods were found when Hbmass > 905 g and BV > 6193 ml.

Conclusion: Assessment of SpCO by Rad-57 resulted in considerably less precise determinations of Hbmass and BV, especially for high values. Thus, non-invasive assessment of Hbmass and BV cannot be recommended for scientific purposes, but may nonetheless be useful in clinical settings.

Keywords: CO pulse oximeter; CO rebreathing; blood gas analysis; carboxyhemoglobin; transcutaneous blood gas monitoring.

Publication types

  • Validation Study

MeSH terms

  • Administration, Inhalation
  • Adult
  • Biomarkers / blood
  • Blood Volume Determination / instrumentation*
  • Blood Volume*
  • Carbon Monoxide / administration & dosage
  • Carbon Monoxide / blood*
  • Equipment Design
  • Feasibility Studies
  • Female
  • Hemoglobins / metabolism*
  • Humans
  • Linear Models
  • Male
  • Oximetry / instrumentation*
  • Predictive Value of Tests
  • Reproducibility of Results
  • Young Adult

Substances

  • Biomarkers
  • Hemoglobins
  • Carbon Monoxide