Two-dimensional shear-wave elastography: a new method comparable to acoustic radiation force impulse imaging?

Eur J Gastroenterol Hepatol. 2017 Jun;29(6):723-729. doi: 10.1097/MEG.0000000000000846.

Abstract

Objectives: Two-dimensional shear-wave elastography (2D-SWE) is an ultrasound-based technique for the noninvasive assessment of tissue stiffness. In contrast to the well-established point-shear-wave elastography (pSWE) method acoustic radiation force impulse imaging, there is little evidence on the performance and usefulness of 2D-SWE in the assessment of liver stiffness. Thus, the aim of our study was to compare 2D-SWE versus pSWE.

Materials and methods: 2D-SWE and pSWE were performed in 20 cirrhotic patients, 20 healthy individuals and an elasticity phantom. Stiffness values, examination time and number of measurements were compared. For 2D-SWE, the influence of size of the region of interest (ROI) was assessed.

Results: Elastography values in healthy individuals were slightly higher for 2D-SWE versus pSWE (1.4 m/s, range: 1.21-1.68 vs. 1.23 m/s, range: 1.07-1.39). In cirrhotic patients, there were no significant differences (3.06 m/s, range: 1.83-5.35 vs. 3 m/s, range: 1.67-4.37 m/s). Examination times were significantly longer for 2D-SWE in both patient groups (mean values for healthy/cirrhotic patients: 129.6/157.1 vs. 75/71.6 s). For 2D-SWE, variation of ROI size (5, 10, 20 mm) produced comparable results. After eight measurements, 90% of cirrhotic patients showed less than 5% of deviation from the results after the gold standard of 10 measurements; for healthy individuals, this was observed after six measurements.

Conclusion: 2D-SWE seems to be comparable to pSWE (acoustic radiation force impulse-imaging) in cirrhotic patients, with slightly higher values in healthy individuals. 2D-SWE measurements require considerably more time. For 2D-SWE, ROI size seems to be of minor importance; multiple measurements should be obtained as single measurements differ. These preliminary results should be confirmed in larger patient collectives with histology as the reference standard.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Case-Control Studies
  • Elasticity Imaging Techniques / instrumentation
  • Elasticity Imaging Techniques / methods*
  • Female
  • Humans
  • Liver / diagnostic imaging*
  • Liver / pathology
  • Liver Cirrhosis / diagnostic imaging*
  • Liver Cirrhosis / pathology
  • Male
  • Middle Aged
  • Phantoms, Imaging
  • Predictive Value of Tests
  • Reproducibility of Results
  • Time Factors
  • Workflow
  • Young Adult