Outcomes of scleral buckling using chandelier endoillumination

Acta Ophthalmol. 2017 Sep;95(6):591-594. doi: 10.1111/aos.13326. Epub 2016 Dec 14.

Abstract

Purpose: To report the results and complications of scleral buckling for the treatment of rhegmatogenous retinal detachment (RRD) using 25-gauge chandelier endoillumination.

Methods: A total of 61 patients (61 eyes) with RRD were treated with scleral buckling. For the sclera buckling procedure, a 25-gauge chandelier was inserted through the pars plana for intra-ocular illumination, and retinal tears were identified and treated with episcleral cryotherapy under surgical microscope. On postoperative days 1, 3 and 7, the intra-ocular pressure was measured by a non-contact tonometer. On postoperative months 1 and 3, ultrasound biomicroscopy was used to examine the pars plana incision.

Results: In the surgical procedure, there was no lenticular or retinal damage due to the chandelier insertion. There was no conjunctival bleb formation at pars plana incision and no incidence of endophthalmitis after surgery. The mean intra-ocular pressure was 15.74 ± 2.98, 15.83 ± 2.76 and 16.14 ± 2.52 mmHg on postoperative days 1, 3 and 7, respectively. The one-time retinal reattachment rate was 93.4%. No visible vitreous incarceration was found in the incision of the pars plana.

Conclusion: There was no complication found due to the chandelier insertion in early postoperative period. Chandelier endoillumination is a feasible method for retinal visualization under surgical microscope during scleral buckling.

Keywords: chandelier; endoillumination; rhegmatogenous retinal detachment; scleral buckling; surgical microscope; ultrasound biomicroscopy (UBM).

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Child
  • Equipment Design
  • Female
  • Follow-Up Studies
  • Humans
  • Light*
  • Male
  • Microscopy, Acoustic
  • Middle Aged
  • Ophthalmoscopes*
  • Retina / diagnostic imaging*
  • Retina / surgery
  • Retinal Detachment / diagnosis
  • Retinal Detachment / surgery*
  • Retrospective Studies
  • Scleral Buckling / instrumentation*
  • Surgery, Computer-Assisted / instrumentation*
  • Time Factors
  • Treatment Outcome
  • Visual Acuity
  • Young Adult