Dual mobility cups in revision total hip arthroplasty

Int Orthop. 2017 Mar;41(3):535-542. doi: 10.1007/s00264-016-3363-4. Epub 2016 Dec 8.

Abstract

Aim: The purpose of our study was to evaluate the dislocation and aseptic loosening rates of a dual mobility construct in revision total hip arthroplasty (THA).

Methods: Three hundred thirty-four revision THAs performed between 2006 and 2011, using a dual mobility cup, were included in this retrospective study. The indications for revision were aseptic loosening (70%), infection (21%) and recurrent instability (9%). The minimum follow-up was five years (mean 7 ± 2 years, maximum 10 years).

Results: At the latest follow-up, 11 episodes of dislocation occurred, seven of which were recurrent (1.3% for aseptic loosening, 2.8% for infection and 3.4% for recurrent dislocation). Ten cases of aseptic loosening occurred; in seven of them, a cemented cup into a reinforcement ring had been used (OR = 14, p = 0.0001).

Conclusion: This study provided evidence of the advantages of dual mobility cups in all revision THA indications.

Keywords: Aseptic loosening; Dislocation; Dual mobility; Revision; Total hip arthroplasty.

MeSH terms

  • Aged
  • Arthroplasty, Replacement, Hip / adverse effects*
  • Female
  • Follow-Up Studies
  • Hip Dislocation / surgery*
  • Hip Prosthesis / adverse effects*
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design / adverse effects
  • Prosthesis Failure
  • Reoperation / adverse effects*
  • Reoperation / methods
  • Retrospective Studies
  • Time Factors